Non- contact infrared thermometer ST-380

Compact, rugged and easy to use. Just aim and push the button, read current surface temperatures in less than a second. Safely measure surface temperatures of hot hazardous or  hard–to-reach objects without contact.
How it works
Infrared thermometer measures the surface temperature of an object. The unit’s optics sense emitted, reflected and transmitted energy which is collect and focused  onto a detector. The unit’s electronics translate the information into a temperature reading which is displayed on the unit. For increased ease and accuracy the laser pointer makes aiming even  more precise.


Cautions
Infrared thermometer should be protected for the following :

  • EMF/electro-magnetic fields/ from arc welders, induction heaters
  • Thermal shock /cause by large or abrupt ambient temperature changes allow 30 minutes for unit to stabilize  before use/
  • Do not leave the unit on or near objects of high temperature .

Warning
Do not point laser directly at eye or indirectly off reflective surfaces.

  1. When take measurement, point thermometer toward the object to be measured and hold the yellow trigger. The object under test should be larger than spot size calculated by the field of view diagram.
  2. Distance and spot size : as the distance from the object increase the spot size of measuring ares becomes larger.
  3. Field of view : make sure the target  is larger than the unit’s spot size. The smaller the target the closer measure distance. When accuracy is critical, make sure  the target is at least twice as large as the spot size .
  4. Emissivity: most organic and painted  or oxidized  surfaces have an emissivity  of 0.95. Inaccurate readings will result from measuring shiny or polished metal surfaces. To compensate, cover the surface to be measured with masking tape or flat black paint. Measure the tape or painted surface when the tape or painted reach the same temperature as the material underneath.

Quick start instruction

  1.  Press battery door clip, install battery correctly. Pull the trigger LCD display reading and battery icon. Release trigger.
  • Laser on icon
  • Scanning  icon
  • Battery power icon
  • F unit
  • C unit
  • Measuring reading
  • Maximum measuring reading
  • Data hold icon
  • Max icon
  1. Locating a hot spot : to find a hot spot aim the thermometer outside  the area of interest, then scan  across  with up and down motions untilk you locate the hot spot. /please turn on the laser to for accurate measuring/
  2. Diagram description
  • Trigger : when pulling the trigger LCD display reading  with SCAN icon. Release the trigger, display reading hold icon.
  • Battery door : when replace battery door please using the finger indents to pull open the battery door.
  • Laser on/off button
  • Celsius / Fahrenheit switch button
  1. Maintenance
  • Lens cleaning: blow off close particles using clean compressed air. Gently brush remaining debris away with a moist cotton cloth.
  • Case cleaning: clean the case with a damp sponge/ cloth a mild soap.

Note :

  1. Do not use solvent to clean lens.
  2. Do not submerge the unit in water.

Specifications

 

Temperature range

-32 to 380 0C /-25.6 to 716 F/

accuracy

25 0 C to 380 0 C+/- 20 C
+/-2%
00 C to 250 C +/-20 C or +/-2%
-320 C to 00 C +/-30 C or +/-2% whichever is greater

Repeatability

1% of reading or 10 C

Response time

500mSec 95% response

Spectral response

5-14um

Emissivity

Pre-set 0.95

Ambient operating range

0 to 50 0 C/32 to 120 0 F/

Relative humidity

10-95%RH noncodensing @ upto 300 C

Storage temperature

-0 to 500 C /32 to 120 0 F/ without battery

Weigth / Dimensions

145g Alkaline or Ni Cd Battery

Battery life /Alkaline/

Laser Models : 16 hrs

Distance to spot size

12:1

Finland Sevastopol joomla